Installation Procedure - Philips Smart-hopping 2.0 Installation Manual

Access point 1.4 ghz
Table of Contents

Advertisement

Installation Procedure

To install the Smart-hopping 2.0 Access Point 1.4 GHz:
1. Mount the Access Point where they can communicate with Patient Monitors. Note
these guidelines when installing the Smart-hopping Access Point:
Caution The coaxial cable bundle cable length may vary from 69 - 79 feet. When planning remote
antenna placement, ensure the maximum length between the Access Point and the
remote antenna is 69 feet.
1. If you have installed Remote Antennas, connect each Smart-hopping Access Point to
its installed Remote Antenna(s) using the supplied, unmodified 74 ft. (22.6m) Coaxial
and UTP cable bundle(s).
Caution Ensure you do not kink the RA Coax-and-UTP-cable-bundle during installation. You must
maintain a minimum 2.5 inch (64-mm) bend radius for the RA Coaxial and UTP cable
bundle throughout the installation.
a) Be sure to label the UTP cable bundles and the Remote Antennas themselves as RA
b) Route a Category 5e (or higher) RJ-45 cable between each Smart-hopping Access
Smart-hopping 2.0 Access Point 1.4 GHz Installation Guide
You may mount each Smart-hopping Access Point as follows:
- To a wall (fixed mount) as described on page 2-3 and page 2-7
- Below a ceiling (flush mount) as described on page 2-12
- Below a ceiling (fixed mount) as described on page 2-20 and page 2-29
Mounting instructions for Smart-hopping Remote Antenna are available in the
Smart-hopping Remote Antenna Installation Guide document, which is available on
the Philips InCenter web site.
Use a category 5e (or better) Unshielded Twisted Pair cable to connect each Smart-
hopping Access Point to the Smart-hopping infrastructure.
The total length of UTP cable from the Smart-hopping access point to the Smart-
hopping Synchronization Unit to the network switch cannot exceed 328 ft.
(100 m).
Use only the supplied, unmodified 74 ft. (22.6m) Coax and UTP cable bundle when
connecting Remote Antennas to the Access Point.
1 and RA 2 corresponding to the cable connections you made when mounting the
Access Point.
Point and the equipment closet in which the Smart-hopping infrastructure devices
are installed.
DRAFT - 03 JUN 2022 13:32
Installation Procedure
2-41