Miscellaneous
Fig. 2
Replacement of mains cable/cable winder
•
Remove the side covers and the cover with the cassette compartment.
•
Remove the four (two on each side) screws that fixes the cable winder assembly
(pos. 1 / Fig. 2).
CAUTION
•
Pull out the cable winder assembly by approximately 10 cm.
•
Disconnect the main switch element by turning the red lock counterclockwise
(pos. 1 / Fig. 3).
•
Disconnect the wires at U9/U10 and ground connection (pos. 2 / Fig. 3).
Siemens AG
Medical Solutions
1
MOB00618
Sharp edge!
The brake handle could be damaged.
Lock the brake handle, otherwise it will be damaged by the sharp
upper edge of the cable winder assembly.
SPR8-220.061.03
Rev. 03
2
1
Fig. 3
Page 3 of 8
09.03
CS PS 24
8 - 3
MOB00624
8
MOBILETT Plus HP