Replacement Of Battery Indicator D210 - Siemens MOBILETT Plus HP Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

7 - 8
Remove the board including the aluminium plate.
NOTICE
Replace the board.

Replacement of battery indicator D210

Remove the right cover and the cover with the cassette compartment.
Remove the two screws (pos. 1 / Fig. 15).
Disconnect K203.
Replace the battery indicator.
MOBILETT Plus HP
Battery circuit and power conversion unit
If the MOBILETT Plus HP is equipped with the DAP measuring
system (optional):
The power supply for the DAP measuring system must be discon-
nected during this procedure. Make sure to remember how the
cables were connected, and if needed use the "Installation
Instructions, DAP measuring system, SPR8-220.031.04..." for
additional help.
SPR8-220.061.03
Rev. 03
Page 8 of 8
09.03
CS PS 24
7
Siemens AG
Medical Solutions

Hide quick links: