Preparation For Use During Surgery Or For Emergency Cases; Preparation Of A Backup System; Power Off/On In The Case Of System Failure - Toshiba Aplio 500 TUS-A500 Operation Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

NOTE: 4. Directly after the system power is turned ON, the fan noise is initially loud
but reduces after a short while.
5. Rarely, the message shown below may be displayed when an attempt is
made to start the system from the Standby status. In such cases, select
[OK] to close the message dialog. Then shut down the system and start it
up again.
Displayed message: [The system failed to start. Please restart the
system. Error code: XXX]
6. After the main power has been turned OFF on the power panel, wait for at
least 15 seconds before turning the power ON again.
If the main power switch is turned ON immediately after the power has
been turned OFF, the system may malfunction.
9.5
Preparation for Use During Surgery or for Emergency
Cases
WARNING:

9.5.1 Preparation of a backup system

Prepare a backup system when an urgent examination is to be performed (for example,
examination during emergency surgery).

9.5.2 Power OFF/ON in the case of system failure

When any of the following abnormalities occurs in the system, recovery may be possible
by turning the system power OFF/ON.
 An error message is displayed and does not disappear.
 The screen display is abnormal.
 System operations are disabled.
If the system is still abnormal even after power OFF/ON, use the backup system.
NOTE:
When turning ON the power, confirm that the [STANDBY] LED is lit and then
press
Prepare a backup system when an urgent examination is to
be performed, for example during an operation or for an
emergency case. If a severe problem should occur with the
system, the normal operation of the system may not be
recovered by turning the power OFF/ON and the examination
may not be able to continue.
.
No. 2B771-004EN*M
9-12
*