To reinstall a foreline gauge

To reinstall a foreline gauge
Materials needed:
Foreline gauge assembly (G1099-60545)
Screwdriver, flat-blade, large (8730-0002)
1 Connect a new foreline gauge assembly to the foreline hose.
2 Tighten the hose clamp.
3 Reconnect the foreline gauge cable to the foreline gauge.
4 Reconnect the foreline gauge assembly to the diffusion pump outlet.
5 Reconnect the MSD to the GC (page 120).
6 If necessary, slide the foreline pump back under the vacuum manifold.
The foreline pump may be located on the floor, on the lab bench next to or behind
the MSD, or under the vacuum manifold at the back of the MSD.
7 Pump down the MSD (page 58).
128
Maintaining the MSD