and pulse detection tone. The sounds can be turned off (muted) on a per patient
basis in the Telemetry Setup window.
Sound 1
(beep)
Sound 1
(beep)
Sound 2
(low pitch
tone)
Service
Two additional sounds that you may hear occur during the labelling of
Sounds
transceiver devices within the IntelliVue Clinical Network. The volume of these
sounds is not adjustable, and cannot be turned off. See "Label Assignment for
Replacement Transceiver" on page 11-2.
Draft - 1 Aug 08
Auditory
Signal
(sound)
Function
Spot Check
Check
Button
Pulse
Detection
Tone
Service Sound
beep beep repeated every
3 seconds, and all visual
indicators flashing
Fast beep beep and
alternate Leads Off
indicators flashing
How Used
Announces a successful SpO
measurement.
Confirms contact with the Information Center.
Indicates pulse during SpO
measurement.
Description
Unlabeled Device.
Equipment Label needs to be assigned at
Information Center. No monitoring.
Unacknowledged Labeled Device.
Equipment label has been assigned at
Information Center; press the Check button to
complete label assignment. No monitoring.
Transceiver Controls
Audible Tones
Spot Check
2
Spot Check
2
3-13