Philips IntelliVue TRx Instructions For Use Manual page 216

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Pairing Non-networked Devices
The following table summarizes alarm behavior when a transceiver is paired
with an IntelliVue bedside monitor without network connection. For detailed
information, see the IntelliVue Patient Monitor Instructions for Use and the
IntelliVue Information Center Instructions for Use.
Where Initiated
Bedside
Information Center
Bedside
Information Center
Bedside
Information Center
Bedside
Information Center
10-20
Pairing Monitoring Devices
Draft - 1 Aug 08
Effect at Bedside
Silences bedside alarms.
No effect
Suspend/Pause
Suspends/Pauses bedside and
telemetry alarms.
No effect
Suspend/Pause not available at the
Information Center
Standby
Bedside put into standby state.
N/A
Standby not available at the
Information Center.
Resume Monitoring
Bedside resumes monitoring
N/A
Resume Monitoring not available at
the Information Center.
Effect at Information Center
Silence
No effect.
Silences telemetry alarms.
Suspends/Pauses telemetry alarms.
All Arrhythmia alarms are off.
No effect.
Suspend/Pause not available at the
Information Center
Measurements from bedside
disappear. Telemetry monitoring
continues.
N/A
Standby not available at the
Information Center.
Measurements from the bedside
reappear, for example NBP and/or
Temp, in the Patient Window.
N/A
Resume Monitoring not available at
the Information Center.