Philips IntelliVue TRx Instructions For Use Manual page 253

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Draft - 1 Aug 08
Parameter
Gain Accuracy
Noise Referred to ECG
Input (Peak-to-Peak)
Lead Wires
Time to baseline from
Defibrillator
Pacer Rejection
Performance
(Pace pulses with no
tails).
EMC Performance
Limits, radiated
immunity
Specification
o
o
+/- 5% at 25
C (77
F)
AAMI: 30 μV (as per AAMI EC 13)
3, 5 or 6-wire leadset compatible with IntelliVue
Patient Monitor, AAMI/IEC color codes
AAMI: 5 s max (until ECG wave is on display
but not yet centered, monitoring bandwidth)
1
Positive pacers
Amplitude
+2 to +700 mV
+2 to +500 mV
+2 to +400 mV
1
Negative pacers
Amplitude
-2 to -700 mV
-2 to -500 mV
-2 to -400 mV
1
Philips does not claim, verify, or validate
support for all available pacemakers.
Meets Essential Performance, but may observe
some waveform disturbance over 181 to 202 MHz
@ >1.8 V/m and 203 to 213 MHz @ > 1.0 V/m
Safety Standards & Specifications
Measurement Specifications
Width
0.1, 0.2, 0.5 and 1.0 ms
1.5 ms
2 ms
Width
0.1, 0.2, 0.5 and 1.0 ms
1.5 ms
2 ms
12-17