62
The code chip is from
an expired lot of test
strips. Ensure the code
chip number matches
the code number on the
test strip container.
Make sure the time and
date in the meter are
correct.
Blood or control solution
was applied to the test
strip before the flashing
drop appeared on the
display. Discard the test
strip and repeat the
test.
An electronic error occurred
or, in rare cases, a used test
strip was removed and
reinserted. Turn the meter
off and on, or take the
battery out for a few
seconds and reinsert it.
Perform a blood glucose or
control test. If the problem
persists, call the Accu-Chek
Enquiry Line.
The temperature is above
or below the proper range
for the meter (6°C to 44°C).
Move to an area between
6°C and 44°C, wait
five minutes and repeat the
test. Do not artificially heat
or cool the meter.

Hide quick links: