Control Tests

Performing a Control Test

You.need.the.meter,.a.test.strip,.and.control.solution.Level.1.or.Level.2.
1
Check.the.use.by.date.on.
Insert.the.test.strip.into.
the.test.strip.container..
the.meter.in.the.direction.
Do.not.use.test.strips.
of.the.arrows.
past.the.use.by.date.
Place.the.meter.on.a.flat.
surface.
5
When.you.see. .flash,.
there.is.enough.control.
solution.in.the.test.strip.
Touch.the.drop.to.the.
front edge.of.the.yellow.
window.of.the.test.strip.
until.you.see. .flash.
Do not put control
solution on top of the
test strip.
28
51746_07022590001.indb 28
2
3
or
Select.the.control.solution.
to.test..You.will.enter.the.
level.later.in.the.test.
6
Wipe.the.tip.of.the.bottle.
with.a.tissue..Cap.the.
bottle.tightly.
4
Remove.the.cap.from.the.
control.solution.bottle..
Wipe.the.tip.of.the.bottle.
with.a.tissue.
Squeeze.the.bottle.until.a.
tiny.drop.forms.at.the.tip.
7/11/13 1:27 PM