Accu-Chek Performa Owner's Booklet page 65

Hide thumbs Also See for Performa:
Table of Contents

Advertisement

Available languages

Available languages

306_33481_PerfManLAapacEN.qxp
The code chip is from
an expired lot of test
strips. Ensure the code
chip number matches
the code number on the
test strip container.
Make sure the time and
date in the meter are
correct.
Blood or control solution
was applied to the test
strip before the flashing
drop appeared on the
display. Discard the test
strip and repeat the
test.
3/9/07
8:42 AM
Page 63
An electronic error occurred
or, in rare cases, a used test
strip was removed and
reinserted. Turn the meter off
and on, or take the battery
out for a few seconds and
reinsert it. Perform a blood
glucose or control test. If the
problem persists, call Roche
Diagnostics.
The temperature is above or
below the proper range for
the meter (6°C to 44°C). Move
to an area between 6°C and
44°C, wait five minutes and
repeat the test. Do not
artificially heat or cool the
meter.
63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents