Calibration
6
an inconspicuous area on the aluminum casting, for the probe to make direct
contact with the metal.
1. Connect the AC power cable to the treadmill.
2. Use a digital multimeter to check all the metal surfaces of the equipment as
illustrated below. Make adjustments for any resistance in the test leads.
3. If the measurements are significantly out of range, check for breaks in the
power cord or in the internal connections within the unit.
4. Make sure all screws on the handles are well secured.
To activate the Series 2000 Treadmill self-calibration feature, do the following:
1. Connect the AC power cable on the treadmill to the wall outlet.
WARNING
POWER CABLES — Route the AC Power cable away
from moving parts.
An AC power cable caught in moving parts could cause
injury to the user and equipment.
2. Turn the power switch on the treadmill to On.
NOTE
Make sure you unlock the emergency stop switch.
3. Remove the plate covering the service port on the rear panel of the treadmill.
4. Remove controlling device (CASE, MTC-1) cable or turn off the controlling
device.
Series 2000 Treadmill
<0.1Ω