Philips CSM 66 III Preinstallation Manual page 28

Table of Contents

Advertisement

SUBJECT
SUBJECT
MAINS VOLTAGE SUPPLY
- mains supply
- voltage variation
- power consumption
- mains connection
- frequency
FUSING/EARTHING
- resistance
MAINTENANCE AREA
STORAGE AND TRANSPORT
- storage needed?
- environmental conditions for stor-
age and transport
. temperature
. relative humidity
- transport equipment available?
- unpacking tools available?
22
Pre-installation Conditions
Pre-installation
Conditions Check
REQUIREMENTS
REQUIREMENTS
200 to 240 V
±10% (inlc. transients)
2.0 kVA (nominal)
4.0 kVA (peak)
1-phase with earth via
2
2.5 mm
conductors
50 Hz or 60 Hz ±2%
< 0.1 ohm
-25 to +55 ˚C (< +70 ˚C
for 24 hours)
5 to 90% (no condensa-
tion)
Pre-installation Manual CSM 66/84/84V III
PA 1306/40 /50 /60 /61
Check List
List
ACTUAL (MEASURED)
ACTUAL
(MEASURED) VALUE
yes/no
yes/no
yes/no
VALUE
REMARKS
REMARKS
Sect. 4.2.1
Sect. 4.2.2
Chap. 5
Chap. 6
Sect. 2.3.1
5322 871 63303
96.00

Advertisement

Table of Contents
loading