GE ApexPro Service Manual page 70

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

3-14
Installation: Mounting
The side mounting brackets are designed to be used in three locations along the sides
of the ApexPro Telemetry Server: end, center, and offset. Choice of mounting
option is determined by rack requirements.
1.
Select a position in the rack for the ApexPro Telemetry Server, according to
rack needs: end, center, and offset.
2.
Remove the screws from the center sides of the ApexPro Telemetry Server,
required for the specific installation.
3.
Position the side brackets so the holes in the brackets align with the center side
holes in the ApexPro Telemetry Server as required for the installation.
CENTER
END
4.
Secure the side brackets to the ApexPro Telemetry Server with the screws
removed in step 2.
5.
For the different rack mounting options:
a.
If you are using either the end, or offset rack mounting option, install the
shelf mount according to
the off-balanced weight of the ApexPro Telemetry Server.
127(
The brackets used for handle installation can be used instead of the side
brackets for the end rack mounting option.
ApexPro Telemetry Server
2001989-178
OFFSET
Holes in bracket
clear screws
"Rack Shelf Mounting"
on page 3-16, to support
174A, 175A, 183A
157A
Revision C