GE ApexPro Service Manual page 185

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Revision C
Parts Lists, Drawings and Replacement: Disassembly/Assembly
4.
Loosen the thumbscrew securing the power supply in the chassis.
5.
Grasp the handle on the back of the power module. Pull the power supply out of
the ApexPro Telemetry Server.
6.
Place the power module in a safe place and allow it to cool. Ensure that no one
touches it with bare hands it until it has cooled to a safe temperature.
7.
Reverse the steps in this procedure to install the power modules.
8.
If applicable, perform the
ApexPro Telemetry Server
2001989-178
"Electrical Safety Tests"
on page 4-10.
166A
165A
6-51

Advertisement