5 - 22
Potentiometer
Fig. 7

Checking the programmed values

NOTE
Prerequisites
T
The
w
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the Adjustment, Calibrations menu.
Click "Yes" to answer the query.
In the Combo box "Function Group:", select "Diamentor".
In the window "Available Functions" , select the line "Start Diamentor Test" and click on
"Execute" .
Close the status window by clicking on the Ok button.
Respond to the query "Do you want to reset the unit" by clicking "No".
Press the -0- key (reset key for fluoroscopic timer).
The tested
This value must match the programmed CONSTANT (Qt*Kg) value. Tolerance: + 5 %.
By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is dis-
played again on the monitor.
SIREMOBIL Iso-C
System Manual
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.
value is already programmed. Refer to Programming.
T
value is displayed on the monitor of the SIREMOBIL Iso-C.
w
Adjustments / Programs
SPR2-230.061.01
Rev. 13
12.04
Page 22 of 30
CS PS 24
5
Siemens AG
Medical Solutions