Setting The Dose Rate For Zoom Format - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Adjustments / Programs
Use the
line).
Apply fluoro briefly until the generator finishes regulating the automatic fluoro adjust-
ment.
Select KV STOP at the system operating console.
Switch on fluoro.
Set the dose rate with the [+kV] and [-kV] keys. (Dose rate * grid correction factor).
The resulting KV value should be between 70 to 80 kV. If necessary, change the prefiltra-
tion and set the dose rate again using keys [+kV] or [-kV]. Refer to the table.
Switch off fluoro.
Remove the dose measurement chamber from the I.I. input screen.
Switch on fluoro.
Press the [>0<] at the system operating console. This starts the automatic adjustment
procedure for the TV- iris collimator.
Wait until the message "done" appears in the "AUTO REGULATION" line.
The "BRIGHTNESS DEVIATION" value is shown in the "BRIGHTNESS DEVIATION"
line and should be 0 + 3 .
Switch off fluoro.

Setting the dose rate for zoom format

refer also to notice on previous page
KV-STOP remains selected.
Center the dose measurement chamber on the I.I. input screen and attach.
Select the I.I. zoom format (refer to "I.I. ZOOM STATUS"). The KV values remain the
same. The mA values, however, are automatically increased by a factor of 1.41.
Use the [mA+] and [mA-] keys to set the dose rate for the zoom format (Dose rate *grid
correction factor).
Press the [>0<] key at the system operating console. This starts the automatic adjust-
ment procedure for the TV-iris collimator.
Wait until the message "done" appears in the "AUTO REGULATION" line.
The "BRIGHTNESS DEVIATION" value is shown in the "BRIGHTNESS DEVIATION"
line and should be 0 + 3. Refer also to notice on previous page
Switch off fluoro.
Select the next level of dose rate (LOW / MID / HIGH) and repeat the dose rate setting for
the survey and zoom format for this level.
Siemens AG
Medical Solutions
key to select the dose rate (LOW / MID / HIGH, shown in the "DOSERATE"
SPR2-230.061.01
Rev. 13
*
Page 7 of 30
12.04
CS PS 24
5 - 7
*
SIREMOBIL Iso-C
System Manual
5

Advertisement

Table of Contents
loading

Table of Contents