Replacing boards / replacing components
Fig. 13

SIMOMED HM monitor

NOTE
Switch the system off.
Replace the monitor.
Perform the monitor settings as described in the Startup Instructions, Chapter 8.
Perform the IQ quick test.

3D reconstruction system electronics

NOTE
The 3D reconstruction system electronics (PC boards D200 incl. MCB300 and D10) is
attached in front of PC board D1 in the basic unit (Fig. 13).
Switch the system off.
Unplug all connectors.
Replace the module.
Reconnect all connectors.
Fasten the shielded cables in the shield connectors again, ensuring good ground con-
tact of the shielding.
If cable ties have been removed, attach the cables with cable ties again.
Perform a download of the control software as described in the Startup Instructions,
Chapter 5.
Perform the mechanical settings as described in the Startup Instructions, Chapter 6.
Perform chapters 10 and 11 of the IQ quick test.