Replacing boards / replacing components
Fig. 13
SIMOMED HM monitor
NOTE
•
Switch the system off.
•
Replace the monitor.
•
Perform the monitor settings as described in the Startup Instructions, Chapter 8.
•
Perform the IQ quick test.
3D reconstruction system electronics
NOTE
•
The 3D reconstruction system electronics (PC boards D200 incl. MCB300 and D10) is
attached in front of PC board D1 in the basic unit (Fig. 13).
•
Switch the system off.
•
Unplug all connectors.
•
Replace the module.
•
Reconnect all connectors.
•
Fasten the shielded cables in the shield connectors again, ensuring good ground con-
tact of the shielding.
•
If cable ties have been removed, attach the cables with cable ties again.
•
Perform a download of the control software as described in the Startup Instructions,
Chapter 5.
•
Perform the mechanical settings as described in the Startup Instructions, Chapter 6.
•
Perform chapters 10 and 11 of the IQ quick test.
Siemens AG
Medical Solutions
Only applies to SIREMOBIL Iso-C 3D.
Only applies to SIREMOBIL Iso-C 3D.
SPR2-230.061.01
Rev. 13
Page 39 of 48
12.04
CS PS 24
6 - 39
6
6
SIREMOBIL Iso-C
System Manual