(1) Preparing for measurement
A) There are nine files in the D:[ccr] folder as Xpress software as indicated below.
Xpress.exe
Callcapt.dll
Comdlg32.ocx
(Xpress.exe is the execution file)
B-1) Set the X-ray generator to the following imaging conditions and align the exposure
position with the imaging unit.
[Imaging conditions]
B-2) Only mAs will be displayed according to the X-ray generator being used. In this case, set
the exposure conditions as follows.
[Imaging conditions]
2 Installation Manual
phase1.sc
phase2.sc
phase3.sc
X-ray tube voltage:
X-ray tube current:
Irradiation time:
Focal distance:
Phototimer:
*If the phototimer is not disabled, an accurate Pre Delay
and Post Delay cannot be set.
X-ray tube voltage:
X-ray tube current:
mAs:
Focal distance:
Phototimer:
*If the phototimer is not disabled, an accurate Pre Delay
and Post Delay cannot be set.
- 111 -
phase1.ini
phase2.ini
phase3.ini
100 kV
50 mA
50 msec
Align with the grid used
Disabled (OFF)
100 kV
50 mA
Set the mAs value between
2.0mAs and 4.0 mAs
Align with the grid used
Disabled (OFF)