bayBIRD Tests
See:
7-5-3 "Diagnostics Window Overview" on page 7-16
features.
Symptoms:
3D triangulation position is off/wrong
V Nav Issues/poor tracking
NOTE:
RF/EMI interference may also cause conditions of these symptoms. See:
on page
1.) Select the bayBIRD Tests checkbox to run all the tests, or select the individual subtests to run only
the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).
Sensor 4 added for DriveBay2+ 3D Tracking System, introduced in R4.
7 - 62
. 2
EV
2-5.
Section 7-5 - Common Service Desktop
LOGIQ E9 S
for a description of the looping and list