7-5-8-1
Digital Receive (cont'd)
Name: Nathan MLA Data Alignment Test
Description: Output TVG of Nathan with system configured for max number of MLAs supported by
system. Each MLA gain is set differently to detect MLA ordering problems. Verifies the alignment (16
identical samples in a row) of IF to GFI.
Run Time: 00:05
If this test failed:
1.) Swap DRX boards to determine if the failure point changes as the boards are moved.
2.) Replace any failed DRX boards.
. 2
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 S
M
ERVICE
ANUAL
7 - 33