Performance Tests
External Mux/Demux
N OT E
It may be necessary to visibly fit the waveform to the mask by adjusting
the Oscilloscope vertical position and horizontal delay to obtain a PASS
on the Mask.
12 Measure the peak pulse amplitude at the mid pulse-width using the
Oscilloscope and verify that this is between 1.896V and 2.844V.
13 Disconnect all test equipment.
3-53