HP 37718A Installation And Verification Manual page 78

Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

N OTE
Performance Tests
PDH/DSn Receiver Equalization
DS1 (1.544 Mb/s)
28 Set the
TRANSMIT
29 Connect Cable Simulator DS1 between the 110Ω OUT and 110Ω IN
ports via two Balanced to Unbalanced Converters (HP 15508B).
30 Press
RESULTS
measurement.
31 Ensure that NO TROUBLE is displayed on the
30 seconds.
32 Press
RUN/STOP
If Cable Simulator DS1 is not available, it is permissible to select DS1-LO
on the
display and connect the 110Ω OUT port direct to the
TRANSMIT
110Ω ΙΝ port. This setup assumes the DS1-LO output signal is within
specification.
3-38
display as shown below.
then
TROUBLE SCAN
to stop the measurement.
to start the
RUN/STOP
display after
RESULTS