CAUTION:
2. Voltage setting for replacing the transformer
tray and TRANS-unit
The switches used to select the output voltage according to the input voltage are
installed on the transformer
tray and the TRANS-unit.
When the switches of the replacement
units (transformer
tray and TRANS-unit)
are not
set correctly, the power supply may not start up correctly, or a malfunction
of the
connection equipment
may result. Be sure to check the settings.
For the transformer
tray
100 to 120 VAC
"115" must be displayed on Sl and S3.
The direction of S2 must be the same as that of Sl
230 VAC
"230" must be displayed on Sl and S3.
The direction of S2 must be the same as that of S3
For the TRANS-unit
100 to 120 VAC
"115" must be displayed.
"115" must be displayed.
The direction must be the same as that of Sl and S2
230 VAC
"230" must be displayed.
"230" must be displayed.
The direction must be the same as that of Sl and S2