Toshiba Aplio SSA-770A Service Manual page 182

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(h)
Click the [Next] button in the Position Adjustment
guide screen shown in (g). Following
the instructions on the Adjust the Range automatically
screen, select the "Range
Adjustment"
icon using the [A], [VI, [-I, and [+] switches and then press the [V] switch to
display the range adjustment
menu.
Use the [-] or [+] switch to select "Execute" and then press [V] to start up the Range
Adjustment
function.
The OSD menu icons are displayed in the order shown below.
When adjustment
is completed,
the display returns to the Screen menu.
Press [v].
Press [A].
Press [v].
Range Adjustment
is executed.
After Range Adjustment
is started up, the OSD menu icons are displayed in the order
shown below. When adjustment
is completed,
the display returns to the Screen menu.
7-15