Siemens Maxum II Manual page 64

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
2540
? I2C Message Too Short on
%4
2541
+ I2C FPGA Queue Full on
%4
2542
? I2C FPGA Write Before
End on %4
2543
? I2C FPGA Write After End
on %4
2544
? I2C FPGA Message Too
Short on %4
2545
? I2C FPGA Invalid Check‐
sum on %4
2546
? I2C FPGA Invalid Size on
%4
2547
!
I2C Heartbeat Timeout Ex‐
pired; Resetting Controller
2549
? I2C Heartbeat Counter Mis‐
match on %4
2550
? I2C Invalid Message Sta‐
tus on %4
2551
? I2C Internal Error; Invalid
Block Structure ID on %4
2552
? I2C Internal Error; Invalid
Block Offset on %4
2553
? I2C Internal Error; Invalid
Block ID on %4
2554
? I2C Internal Error; New
Block Offset Non-Zero on
%4
64
Description
SNECON I2C driver: An I2C message
received by the SNECON is too short to
be valid.
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON on-board
communication error.  Known to hap‐
pen very infrequently with SNECON I2C
FPGA rev 23 or lower (SNECON revision
3.0 hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: A SNECON PIC
was not responding and was reset (SNE‐
CON revision 3.0 hardware only).
SNECON I2C driver: One or more mes‐
sages were lost on the on-board commu‐
nication (SNECON rev 3.0 hardware on‐
ly).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
SNECON I2C driver: SNECON firmware
internal error (SNECON revision 3.0
hardware only).
Action
Contact Customer Support.
Reset the analyzer.
Replace the SNECON.
Reset the analyzer.
Replace the SNECON.
Reset the analyzer.
Replace the SNECON.
Reset the analyzer.
Replace the SNECON.
Ignore the alarm. If occurring frequent‐
ly, replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Reset the analyzer.
Reload the SNECON OS software.
Replace the SNECON.
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001

Hide quick links:

Advertisement

Table of Contents
loading