General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
#
Text
676
!
Database: delay limit ex‐
ceeded on stream %3
677
? Next Stream Error
678
!
End of Cycle occurred be‐
fore events completed
679
!
Application is out of service Attempt was made to move (approve) re‐
680
? Print job failed: %3 for Print‐
er: %4
681
!
Application is disabled
682
!
Database: I/O failure: %3
683
? Database: no normal se‐
quence for application: %3
684
? Database: no enabled en‐
tries in sequence
685
? Printer: TCP connection or
queue failed
686
? Printer: TCP Print failed
687
!
Results for cycle %3 lost
due to SNE reset
688
!
Method is corrupt: %3 re‐
load from EZChrom
689
!
Database: Fault on Slave
Application %3
690
!
Database: Slave Applica‐
tion not Complete %3
691
? Database: Warning: %3
692
!
Database: Divide by zero in
%3
693
? Database: I/O warning: %3 System error
694
? Database:value > limit: %3 Limit exceeded. Message should contain
695
? Database value < limit: %3 Limit exceeded. Message should contain
696
? DB: Screen access denied
697
!
DB: Run requested on dis‐
abled program: %3
42
Description
Temperature or pressure wait set in the
method has been exceeded.
Cannot locate next stream in sequence
after a stream step.
System error
sults for an out-of-service application.
Print failure.
Attempt was made to set disabled appli‐
cation to run.
Bad status on AO, DO write.
No active sequence.
Cannot find enabled entry in sequence.
Printer connection failed.
Communication failure with TCP printer.
This alarm marks results as uncertain un‐
til a cycle has been completed after re‐
setting unit.
Method is corrupt.
A fault is being transferred from the slave
to the master to invalidate the master's
results.
Master application is trying to run when
slave is not in hold. The slave must com‐
plete before the master.
System error
Peak measured value is zero during cal‐
ibration for an autocalibration, so margin-
checking cannot occur.
sufficient information.
sufficient information.
System error
System error
Action
Check temp or press controller.
Check wait_delta and maxwait in tem‐
perature or pressure controller table.
Check sequence to ensure that there is
an enabled stream.
Contact Customer Support.
Put application in service.
Check printer.
Enable application.
Check I/O channels.
Check sequences.
Check sequence.
Check address in printer table.
Try printing again.
Automatic repair: No action necessary.
Download method from EZChrom.
Check fault in slave application.
Check cycle length of slave application;
should be shorter than master. Also
could occur if autocalibration sequence
for master is shorter than for the slave.
Contact Customer Support.
Check method.
Contact Customer Support.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
Contact Customer Support.
Contact Customer Support.
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001