Siemens Maxum II Manual page 59

Pd pa ap. general maintenance
Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

#
Text
2229
!
Mezzanine-module ID inva‐
lid
2230
!
Mezzanine-module ID re‐
served
2257
!
Balance Failure FID on %4 FID DPM Detector PIC: A detector bal‐
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001
4.6 Alarm Codes, Descriptions, and Suggested Actions
Description
FID DPM Detector PIC: The mezzanine-
module ID value is not supported by this
DPM board. More specifically, the mod‐
ule value is pointing to a non-initialized
EEPROM region on the DPM. This may
indicate an invalid mezzanine module or
an older revision of the DPM that was
manufactured before the mezzanine
module was defined.
FID DPM Detector PIC: The mezzanine-
module ID value is set to a value re‐
served for future expansion.
ance sequence failed because of a hard‐
ware failure or because of an improper
configuration or sequence of events.
Some possible causes are:
The detector signal is not stable enough
and proper balance could not be ob‐
tained before the maximum number of
iterations was reached.
The A/D or D/A do not work properly.
The detector is not configured for acquis‐
ition or turned off.
SIMULATE_FID is set to 1 (the balance
cannot be done in detector simulation
mode).
The balance is disabled (DISABLE_BAL‐
ANCE_FID is set to 1).
The sampling period was changed dur‐
ing the balance sequence.
General Maintenance and Troubleshooting
Action
Verify the DPM revision level and re‐
place as needed.
Replace the mezzanine module on the
DPM.
Verify the DPM revision level and re‐
place as needed.
Replace the mezzanine module on the
DPM.
Contact Customer Support.
59

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents