1
Program number can be from 1 to 9.
2
When you want to define a new program, enter 1.
As a result, you will delete the previous program.
For modifying the contents of program, enter 0.
3
If pump pressure rises beyond the level set here, the pump automatically stops.
4
If pump pressure falls below the level set here, the pump automatically stops.
5
The set time can be from 0.0 to 600.0.
6
Flow rate valid entries are from 0.00 to 2.499 (L-2100). Flow rate valid entries are
from 0.00 to 9.999 (L-2130).
7
Input the event code. At the set time point, you can stop the pump, generate an
audible signal, markers, timers, etc. When properly interfaced, these signals can
be output to an external system. No display will show on the screen unless an
event is set.
8
Repeat steps 5 to 7 until you complete the program and then press [ESC] .
9
Program generation will be terminated and the monitor screen will reappear.
For confirming the input conditions, check the parameters at each time point by using
keys.