Isocratic System
Connect the inlet tube of eluent to the IN port of the
degasser. Connect the OUT port of the degasser to the
inlet of the pump. Connect the drain valve of the pump to
port 1 of the injection valve of the autosampler.
Connect port 2 of the injection valve to the column inlet.
Connect the column outlet to the inlet tube of the detector
cell. For washing solution, provide connection to port 1 of
the syringe valve of the autosampler through the degasser.
Washing solution
Model L-2400 UV detector
Model L-2300 column oven
Model L-2100 pump
Model L-2100/2130 pump
Degasser
2 - 11
Eluent
Inlet tube
Detector cell
Column
Injection valve
Syringe valve
Drain
Drain valve