10
ADJUSTMENT
To obtain the best performance, periodically calibrate the
unit. Sometimes, only one mode need be calibrated, while
at other times, all modes should be calibrated. When one
mode is calibrated, it must be noted that the other modes
may be affected. When calibrating all modes, perform the
calibration in the specified sequence.
The following calibration required an accurate measuring in-
strument and an insulated adjusting flat blade screwdriver.
If they are not available, contact your dealer. For optimum
adjustment, turn the power on and warm up the scope suffi-
ciently (more than 30 minutes) before starting.
Before calibrating the unit, check the power supply voltage.
TEST EQUIPMENT
REQUIRED
The following instrument or their equivalent should be used
for making adjustment.
Test Equipment
Model
Maker
Frequency Counter
FC-756
KENWOOD
Digital Multimeter
8600A
FLUKE
Sine-Wave Generator
651B
HP
SSVM
VT-106
KENWOOD
Wow-Flutter Calibrator
MWS-670C
MEGURO
Oscilloscope
| CS-6010
KENWOOD
Distortion Meter
885
ShibaSoku
Watt-meter
DPB-1W
YEW
Insulation Meter
SM-5
TOA
Fuction Generator
FG-273
KENWOOD
PREPARATION
FOR ADJUSTMENT
Control Settings
The contro! settings listed below must be used for each
adjustment procedure.
Exceptions to these settings will be noted as they occur.
After completing a adjustment, return the controls to the fol-
lowing settings.
[
NAME OF KNOBS
POSITION
POWER
ON
ocs
3 kHz
MODE
WOW FLUTTER
LEVEL
5 mV~30V
REFERENCE FREQUENCY
CRYSTAL