Switching The Wireless Foot Switch On And Off - Philips 4523 001 21821 Manual

Wireless foot switch
Table of Contents

Advertisement

3 Operation
Wireless Indicator
NOTE
If the battery indicator is red and flashing rapidly, the red wireless indicator light is also switched on.
This is because there is insufficient charge in the battery to maintain the wireless connection and the
wireless foot switch has stopped working. Connect the charger or use the wired foot switch.
3.3

Switching the Wireless Foot Switch On and Off

NOTE
Ensure that the wireless foot switch battery is sufficiently charged before using the foot switch. For
information about battery charge level, see
depleted during a procedure, the foot switch will stop working. In this case, use the wired foot switch.
Alternatively, connect the charger to the foot switch and continue to use it. Take care not to damage
the cable of the charger when moving equipment around the examination room (for example, when
moving carts or beds).
Before using the system, check that the wireless foot switch functions with the system and all the pedals
are working correctly. If identification labels have been used, check that the labels attached to the system
and to the foot switch match. For more information, see
The wireless foot switch may be put in a sterile plastic cover.
Wireless Foot Switch Addendum
Status
Off
Wireless connection is
operational.
Red
There is an error in the wireless
connection. Do not use the foot
switch.
Red,
A critical error has occurred
flashing
(not applicable for the type 1
rapidly
wireless foot switch).
Switching the Wireless Foot Switch On and Off
Action
No action required. The foot switch is ready to use.
Move the wireless foot switch closer to the receiver. The location
of the receiver depends on your system configuration; it is
located either inside the patient table base or in the surgery wall
connection box. Wait for the wireless indicator to go out before
using the wireless foot switch.
If this does not resolve the issue, continue the procedure using the
wired foot switch.
To resolve the issue after completing the procedure, switch the
wireless foot switch off and then on again or perform a cold restart
of the X-ray system. If this does not resolve the issue, continue
using the wired foot switch and contact technical support.
Switch the wireless foot switch off and then on again or disconnect
the charger if it is connected. Wait for the wireless indicator to go
out before using the wireless foot switch.
If this does not resolve the issue, continue the procedure using the
wired foot switch.
To resolve the issue after completing the procedure, perform a
cold restart of the X-ray system. If this does not resolve the
issue, continue using the wired foot switch and contact technical
support.
Wireless Foot Switch Indicators (page
About the Wireless Foot Switch (page
10
9). If the battery is
5).