Toshiba LF620F Instruction Manual page 59

Electromagnetic flowmeter converter
Hide thumbs Also See for LF620F:
Table of Contents

Advertisement

Switch operation
STEP4
STEP5(=END)
Note 1:The meter size display loops as shown below:
2.5mm
0.1inch
Note 2: When the meter size is changed, range unit and count rate will be forcefully changed as
described below. If necessary, change these parameters again.
Range unit
Count rate
Note 3: The exciting frequency setting may become inappropriate for the set value when the
meter size of the detector is changed. If the exciting frequency is the value shown
below when the meter size is changed, the exciting frequency will be forcefully
changed.
Setting meter size
(mm)
2.5 ~ 200
250 ~ 450
500 ~ 900
Buy: www.ValinOnline.com | Phone 844-385-3099 | Email: CustomerService@valin.com
Display example
SIZE
Repeat this operation until 150mm is
obtained.
150
When the desired value is obtained, push
mm
SIZE
Pushing
whether the setting is OK or not.
150
If OK, push
setting, push
mm
SET OK?
Pushing
and exits the setting screen.
ESC
N O
O K
4mm
6mm
0.16inch
0.25inch
m/s
If the count rate goes out of the setting range
when the meter size is changed, the count rate
will be forcefully set to zero.
Set exciting frequency
(inch)
0.1 ~ 8
10 ~ 18
24Hz
20 ~ 36
12Hz, 24Hz
- 58 -
6 F 8 A 0 9 2 8
Description
to set the value temporarily.
shows a message to confirm
O K
. If you need to redo the
N O
.
cancels the setting operation
ESC
15mm
900mm
0.5inch
36inch
Forcefully set exciting
frequency
Not forcefully set
12Hz
6Hz

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lf622f

Table of Contents