2.Cancel sterilization
During the sterilization process, press the "Cancel" button. A cancel sterilization dialog box will
appear. When clicking "confirm to stop", it will return to the "Settings" interface. The sterilization
stop warning dialog box will pop up. When clicking the "Continue Sterilization", it will return to the
sterilization interface.
• Sterilization pause
a) Door opening and loss of power cancel the process of sterilization. If the door is closed or the
power is recovered, a dialog window for sterilization appears on the screen to prompt the user for
action to abort the process or to continue. The temperature displayed in the dialog window is the
real time temperature in the incubator.
b) An alarm will also sound along with the sterilization pause warning dialog box. Click any
position on the screen, the alarm will be canceled while the sterilization stops warning dialog box
stays. Once the incubator resumes to the set temperature, the sterilization pause warning dialog
box will automatically be canceled.
Sterilization stage
heating stage
Diagram of the dry heat sterilization stage
Sterilization cancel interface
Cooling stage
time(hr)