Warning Health hazard!
The surface of the working chamber may be contaminated. Contact with
contaminated cleaning fluid may cause infection. Disinfectants may contain
hazardous substances.
Always observe the safety instructions and hygiene regulations during cleaning and
disinfecting!
a) Wear safety gloves;
b) Wear safety goggles;
c) Wear masks and other oral cavity respiratory protective devices to protect your
mucosa from stimulation;
d) Observe the safety instructions of the disinfectant manufacturer and the health
authority.
Sterilization operation
This series of incubators is equipped with 180°C dry heat sterilization function. The sterilization
method and sterilization effect comply with WS/T 67-2012 medical institution disinfection technical
specifications. It can kill all microorganisms, including bacterial spores, and reach the sterility
assurance level.
The whole process of sterilization can be divided into three stages: heating stage (about 2h),
continuous sterilization stage (about 1.5h) and cooling stage (about 8h). Before sterilizing, reinstall
the sterilized air duct plate and shelves onto the incubator, connect the power supply and turn on the
power switch.
1.The user can enter the sterilization procedure in two ways:
•Directly click the sterilization icon
•Click the setting icon to enter the setting interface, and then click the sterilization icon to enter the
sterilization procedure;
on the main interface to enter the sterilization procedure;
Sterilization confi rming interface