Setting The Analog Outputs - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.8 [2.09] Settings > Inputs/outputs
6.8.2.2

Setting the analog outputs

Procedure
Note
Component-specific setting
An analog output is available for each component. Repeat the procedure described below for
all components, if necessary.
1. Component > Main menu > "2. Settings" > "09. Inputs/outputs" > "2. Analog output"
The parameter display is opened.
2. Set the analog output range: "Analog output range" parameter box.
– Authorize access with Standard PIN, if necessary.
– Select one of the available list entries according to Table 6-15 Analog output range
3. Permit or suppress negative values: "Negative values" parameter box.
If negative measured values have a negative impact on the measured value processing,
select "Suppress".
Negative measured values are then set to 0 or 4 mA at the analog output while the
display still shows the negative measured value.
4. Set the behavior of the analog output with enabled function check: "At function check"
parameter box.
By setting this parameter you determine which value is output by the analog output on a
change to "Function check" operating state. The following values can be output as an
alternative:
5. Set the behavior of the analog output in case of a maintenance alarm or fault: "At
main.alarm/fault" parameter box:
Setting range and setting options correspond to step 4.
See also
Overview (Page 141)
Setting the measured-value format (Page 98)
142
and measuring range limit (Page 141).
Setting
Description
Min
Low limit of the analog output range
Max
High limit of the analog output range
Measured value
Current measured value
Retain
Last measured value before entering function check
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05