Functions
6.7 [2.08] Settings > Calibrations
If the analyzer module determines a tolerance violation during calibration of a gas
component, the corresponding sequence step is ended and the AutoCal process is
continued as planned. But the tolerance violation is entered into the logbook of the device.
If you cancel an AutoCal, the currently executed sequence step is canceled and the entire
sequence is aborted. Any single calibrations that were successfully conducted up to this
point are retained.
See also
Sequence of a calibration/validation (Page 116)
6.7.6.2
Setting the AutoCal/AutoVal sequence
Requirement
The following requirements must be met before you can specify the operating mode or start
an AutoCal/AutoVal:
● Calibration pool entries are available.
● Starting the sequence using the digital inputs: The digital inputs have been configured
accordingly.
132
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05