15 - 4
High-voltage circuit errors
The following events in the high-voltage circuit lead to the generation of an error:
•
HT PEAK
•
HT
•
HTD
•
OVERCURRENT
The unit reacts as follows to an error in the high-voltage circuit:
•
the signals "INV A" and "INV B" are
•
the green "ready" LED on
the Control and display panel is
•
the rotating anode brake is
•
the exposure heating is
•
the preheating is
•
in the Control and display panel an
The error message displayed can be acknowledged by pressing the selection button
"kV +".
If the same error message appears again proceed according to the Service Instructions!
MOBILETT Plus/ Plus E
> ±5 kV outside the selected kV value
>140 kV
>±20 kV
resulting from a high-voltage inverter current of
>290 A.
Register 6
SPR8-215.850.02
Rev. 01
High-voltage circuit
disabled,
turned off,
switched on,
switched off,
switched on, and
error message is shown.