Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
2.4.7. High Level Point (default 100)
The level when the material is at High Calibration Point. The unit is defined in Level
units (open the menu
to see an illustration.)
2.4.8. Level Offset (default 0)
A constant offset that is added to Level to form SV1 (Secondary Value 1). The unit
is defined in Level units.
Level Offset is used to compensate for specific tank configurations.
2.4.9. Sensor Offset (default 0)
Note:
been set manually, the sensor offset value will be reset to 0.
The offset from the Sensor's reference point to the tank's reference point:. If
sensor offset is used, it is a negative value that is subtracted from the values for
the high and low calibration points. The unit is defined in Sensor Units.
Compensates, for example if the sensor head is changed.
2.4.A. Temperature Units
Selects the engineering unit to be displayed with the value representing
temperature.
Hand
programmer
values
2.5. M easuring Limits
2.5.1. Min. Measured Value
The minimum recorded Sensor value, defined in Sensor units.
Open menu
2.5.2. Max. Measured Value
The maximum recorded Sensor value, defined in Sensor units.
Open menu
Page 82
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
1.
Open the menu
for Wet or Dry calibration.
2.
Enter the values for Low Calibration Point, Low Calibration Level,
High Calibration Point, High Level Point, then click on
If a velocity calibration is performed after the sensor offset value has
, then click on
Degrees K
Degrees C
Degrees F
Degrees R
, and click Sensor tab.
, and click Sensor tab
Doc Id: VM11
Active Date: 15/12/2006
, and click on the button
.
7ML19985JB02
,