Appendix F: Profibus Pa Profile Structure; Profibus Level Device Design; Block Model For Recording And Processing Measured Values - Siemens Sitrans Probe LU Instruction Manual

Hide thumbs Also See for Sitrans Probe LU:
Table of Contents

Advertisement

Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
Appendix F: PROFIBUS PA Profile
Structure

PROFIBUS Level Device Design

The device follows the profile block model and is implemented as a Profile 3.0, Rev. 1,
Class B, PA device. Standard profile parameters are used to program the level transducer
block.
Block M odel for recording and processing measured values
The functions of the device are divided into blocks for different areas of responsibility.
They can be parameterized by acyclic data transfer via PDM.
sensor
The device is implemented with one Physical Block (PB), one Level Transducer Block
(LTB), and two Analog Input Function Blocks (AIFB1 and AIFB2).
All data is viewed from the perspective of the DCS or PLC, so information from the sensor
is an input.
Level Transducer Block (LTB)
The Level Transducer Block carries out adjustments to the sensor, such as level
calibration and volume calibration. It supplies the outputs used by either or both of the
AIFBs.
7ML19985JB02
sensor
electronics
temperature
Transducer Block
transducer
temperature
sensor
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
Doc Id: VM11
Active Date: 15/12/2006
Analog Input
Function Block 1
Analog Input
Function Block 2
Page 55

Advertisement

Table of Contents
loading

Table of Contents