Installing the analyzer modules and option module
5.3 Installation of the analyzer modules
5.3
Installation of the analyzer modules
5.3.1
Device expansion by a second analyzer module
If you install a second analyzer module, change the magnetic field frequency of the
OXYMAT 7:
If the device contains two OXYMAT 7 channels, set the magnetic field frequency of the last
analyzer module installed to 10 Hz.
5.3.2
OXYMAT 7
Mounting
1. Isolate the device from power.
2. Remove the housing cover. To do so, unscrew the two cap nuts on the rear panel and
3. Remove the blank plate at the rear of the housing at the corresponding slot. To do so,
4. Fit the analyzer module
5. Fasten the analyzer module
Note
Tightening the screws
Tighten the screws on the rear panel and housing bottom with a torque of
• 2.5 Nm.
28
push the housing cover to the back.
loosen the four screws and remove the blank plate.
housing bottom.
②
into the rack-mounted housing
②
with four screws to the rear panel and two screws to the
①
.
Operating Instructions, 01/2013, A5E31874006-01
Rack-mounted device