Electromagnetic fields (EMF)

This Philips device complies with all applicable standards and regulations regarding exposure to electromagnetic fields.

Technical information

Vacuum level
Cycle speed
Motor unit rated input
Mains adapter rated input
Mains adapter rated output
Adapter type number
Power port
Operating conditions
Storage and transportation
conditions
Operating pressure
Net weight
External dimensions
Device classification
Materials
Service life
Mode of operation
Ingress protection
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic compatibility (EMC)
as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables. Only use this Philips Avent
single/double electric breast pump Plus in combination with the adapter supplied.
Single electric breast pump
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbarIntermittent vacuum
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
5 V d.c. / 1.1 A
External mains adapter 100-240 V a.c.
50/60 Hz
400 mA
5 V d.c. / 1.1 A
S009AHz050yyyy The letters "yyyy" represent the output current from 0100
(1000 mA) to 0180 (1800 mA), increments in steps of 100 mA. The letter "z" denotes
type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F.
Micro-USB
Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Relative humidity: 15 to 90 % (non-condensing)
Temperature: -25 °C to 70 °C (-13 °F to 158 °F)
Relative humidity: up to 90 % (non-condensing)
700-1060 hPa (<3000 m altitude)
230 g
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Applicable classifications:
Degree of protection against electric shock:
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Double electric breast pump
5 V d.c. / 1.8 A
5 V d.c. / 1.8 A
310 g