Patient-Monitor Type

4-12
Configuration: Service Mode settings
The Patient-Monitor Type selection determines the type of patient
monitor desired, e.g., adult, neonatal or operating room. Different alarms
and parameters are activated for each selection.
CAUTION
Each time the patient-monitor type is changed, the
ADMIT MENU function defaults to STANDARD
configuration. Be aware that some patient alarms and
parameters may be changed.
NOTE
The keypad/remote control is DIDCA programmed for specific
patient monitor types. The error message, "WARNING: REMOTE
MISMATCHED WITH MONITORING MODE" displays if the
patient monitor and keypad/remote control do not match.
1. Enter service mode (See
select PATIENT-MONITOR TYPE. Be sure to read the information
in the Attention box before changing anything.
2. Rotate Trim Knob control to select the type of environment the
patient monitor will be used in.
3. Press Trim Knob control to exit. Your selection displays at the top
left of the screen after the date.
Dash 3000/4000/5000
"Service Mode settings" on page
042A
4-11) and
2000966-542D