GE Dash 3000 Service Manual page 199

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Recommended calibration, electrical safety tests, and checkout procedures
Replacement
Calibration
procedure
tests
DAS assembly
"NBP
calibration" on
page
2000966-542D
Field replaceable units: Recommended checkout
Checkout procedures
"ECG tests" on page
"Respiration tests" on page
8-37.
"Temperature tests" on page
"Cardiac output tests (option)" on page
"Invasive blood pressure tests (option)" on
page
8-27.
"Pulse oximetry tests for GE Ohmeda SPO2
oximeter" on page
you installed a GE Ohmeda DAS assembly.
"Pulse oximetry tests for Masimo SET SPO2"
on page
8-32. Use this procedure if you
installed a Masimo DAS assembly.
"Pulse oximetry tests for Nellcor OxiMax SPO2"
on page
8-33. Use this procedure if you
installed a Nellcor OxiMax DAS assembly.
"Noninvasive blood pressure tests" on page 8-
35.
"Wireless antenna signal strength test (option)"
on page
8-49. Use this procedure if the wireless
option is installed.
"Wireless LAN test (option)" on page
this procedure if the wireless option is installed.
Refer to the CO2 chapter in the patient
monitor's operator manual to checkout the
mainstream End-tidal CO2.
To check out the sidestream End-tidal CO
refer to the documentation that came with the
module.
Dash 3000/4000/5000
8-22.
8-25.
8-26.
8-27.
8-30. Use this procedure if
8-52. Use
,
2
Electrical safety tests
"Electrical safety tests" on page
8-4.
7-53