Appendix C. Factory Tests
C.3
Resetting the XGM868i to Factory Defaults (cont.)
7. Scroll to LCD and press
8. The window now asks for the # of LCD Parameters. Scroll to the desired number (from OFF through 1-4 and
KEY) and press
[Enter].
For a 1-Channel XGM868i, the data for Channel 1 is displayed automatically, and you can skip to step 9. However, for
a 2-Channel meter, the channel data to be displayed must be specified at the following prompt.
9. Scroll to the desired Channel option, as listed in Table 11.
10. For each channel, select the desired Measurement Parameter, as shown in the Programming Manual.
Note: The measurement units that appear in these prompts are those selected in the
The previous two prompts repeat until all of the specified # of LCD Parameters have been set up. When all of the
display parameters have been set up, the meter returns to the
the
key three times.
[Escape]
After leaving the Keypad Program, the XGM868i will reset itself and will begin to display the parameters specified in
this section. If more than one parameter was set up, each of the parameters will be displayed in sequence, with a pause
of several seconds between display changes.
64
[Enter].
Table 11: Channel Options
Option
CH1
CH2
SUM
DIF
AVE
Description
Channel 1
Channel 2
CH1+CH2
CH1-CH2
(CH1+CH2)/2
Global I/O
window. To leave the Keypad Program, press
menu.
GLOBL-SYSTM
DigitalFlow™ XGM868i Service Manual