GE Vivid T8 Basic Service Manual page 234

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Diagnostics/Troubleshooting
7-4-5-6 FRU Status
Element
FRU Status Summary
FRU Status Details
Name
Status
Low Level Warning
High Level Warning
Low Level Errors
High Level Errors
7-24
FRU Status displays a summary of the FRUs being monitored
and the details for each FRU.
To access this page, under System Status, select Temperature
and then select FRU Status.
Figure 7-11. FRU Status
This table shows all the elements available on FRU Status with
descriptions.
Table 7-5:
FRU Status
DESCRIPTION
Number of the FRUs being monitored.
Name of the FRU.
• ETX3 - Represents the ETX 64 transmitter board.
• ETX4 - Represents the ETX 128 transmitter board.
Status of the FRU. Valid values are:
• Normal - Indicates that the FRU is operating within the
allowable range.
• Warning - Indicates that the FRU is operating close to
the limit of the allowable range.
• Error - Indicates that the FRU is operating outside the
allowable range.
Number of low temperature warnings.
Number of high temperature warnings.
Number of low temperature errors.
Number of high temperature errors.