Connecting
6.2 Electrical connections
6.2.5

Connecting wall-mounted device electrically

6.2.5.1

Device-specific safety instructions

Danger due to electrical current
The power cable to be connected at the customer location could be damaged by improper
connection to the device.
• Check whether the existing line voltage corresponds to the line voltage specified on the
• Before starting up the device, you should carry out a visual inspection and PE conductor
• Because part of the cable runs in the device, always adhere to the product safety
NOTICE
Cable glands for the wall-mounted device
• Match the diameters of the cables to the terminal areas of the cable glands.
• After installation, make sure that the cables are secure and that the seals are in the
Note
Terminal ranges of the cable glands
Use cables matched to the terminal ranges of the cable glands:
• Cable gland M16 x 1.5 (plastic): Terminal range 5 to 10 mm
• Cable gland M16 x 1.5 (metal): Terminal range 6 to 10 mm
• Cable gland M20 x 1.5 (plastic/metal): Terminal range 10 to 14 mm
60
WARNING
nameplate of the device or in the technical specifications.
You can also find the required line voltage in the test certificate valid for your country.
test of the cable. To do this, proceed in accordance with the country-specific safety
standards, regulations or other applicable directives.
standard IEC/EN 61010-1 or its national equivalent.
correct position.
Compact Operating Instructions, 05/2018, A5E31805656-06