Note
Alarm window
When you change the hardware of the device by replacing or adding modules or components,
the alarm window "Device configuration changed" is displayed in the LUI during re-
commissioning. To start the data synchronization, confirm the alarm by pressing <ENTER>.
Depending on the specific application, parameter sets may have to be transmitted. The
message window "Transmit parameters" is displayed. Information on the further procedure is
available in the tables below:
→ Case differentiations (Page 53)
→ Case differentiations (Page 53)
→ Case differentiations (Page 53)
→ Case differentiations (Page 53)
→ Case differentiations (Page 53)
3.7.4.3

Case differentiations

Table 3-4
Case
PM
New
New
Table 3-5
Case
PM
Un‐
change
d
Un‐
change
d
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
Plug & Measure - Commissioning
Description
AMx
New
Commissioning, PM without serial num‐
ber
New
Commissioning, each module has a se‐
rial number stored at the factory.
None of the modules has been connec‐
ted to another module before
Plug & Measure - Installation/replacement of an AM
Description
AMx
Re‐
The AM is to be replaced by a module
placed
of the same type. The replacement AM
has a different serial number than the
AM connected previously.
Replacement of the AM electronics, on‐
ly permitted without own serial number
stored in the AM electronics
Re‐
AM electronics installed as "dedicated
placed
spare part". Service event.
3.7 Combined operation
Data synchronization/Action
Automatic data synchronization.
Automatic data synchronization.
Data synchronization/Action
User action. Confirm <ENTER> or re‐
ject <ESC>.
The data is transmitted after confirma‐
tion, except the identification data and
the calibration parameters.
User action. Always confirm <ENTER>.
The data is transmitted after confirma‐
tion, including identification data and
calibration parameters.
Normal startup without message. The
default data of the AM set in the factory
is applied.
Description
53