3.7 Combined operation
CALOMAT 7
Table 3-3
Operating mode
Pressure operation
Suction operation
Retrofitting example: "Pressure operation, serial" for OXYMAT 7/ULTRAMAT 7
When you are replacing the ULTRAMAT 7 with an OXYMAT 7 analyzer module, you must
replace the sample gas restrictor in the sample gas inlet with a clamping screw.
MGE, VGE
Figure 3-15
50
Mounting conditions for the sample gas restrictor: CALOMAT 7
Sample gas restrictor/clamping screw
Operation with one CALOMAT 7
● Pre-assembled clamping screws. No retrofitting or dismantling.
Sample gas inlet (MGE), Reference gas inlet (VGE)
Initial situation, serial pressure operation of OXYMAT 7 (AM1) and ULTRAMAT 7 (AM2)
Remove sample gas restrictor in the sample gas inlet of the replacement module
Install clamping screw in the sample gas inlet of the replacement module
Replace ULTRAMAT 7 with OXYMAT 7
Retrofitting of the sample gas restrictors: Example "Pressure operation, serial"