7 Prepare the system for a run
7.3 Prime buffer inlets and purge System pumps
Step
5
6
7
8
9
80
Action
Connect a 25-30 ml syringe to one of the purge valves of the selected system
pump. Make sure that the syringe fits tightly into the purge connector.
Open the purge valve by turning it counterclockwise about 3/4 of a turn.
Draw 5-10 ml of liquid slowly into the syringe.
Close the purge valve by turning it clockwise. Disconnect the syringe and
discard its contents.
Click the Continue button on the Instrument display to start a pump wash
of the next inlet in the Instruction execution list.
Repeat steps 4-7 until all inlets in the Instruction execution list have been
primed (filled with liquid).
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB